Availability: | |
---|---|
The quality of steam largely determines the effectiveness of the sterilization process, which in turn affects the quality of the final product. According to EN285 and HTM2010, the content of non-condensable gases, dryness, and overheat have a significant impact on the sterilization effect of steam. Therefore, the following requirements are needed for steam:
1. Non-condensable gas: Non-condensable gas can be mixed into pure steam during the preparation process, resulting in a mixture of steam and gas. The volume of non-condensable gas in 100 ml of saturated steam does not exceed 3.5 ml (equivalent to 3.5%, volume fraction);
2.Drying degree: The drying degree is a measure of the total amount of liquid water in steam. The lower the drying degree, the less potential heat released during sterilization. Currently, most drying degree detection methods are approximate. For the sterilization of metal carriers, the drying value should not be less than 0.95; for the sterilization of non-metal carriers, the drying value should not be less than 0.9;
3.Overheat: When pure steam is released to atmospheric pressure, the overheat does not exceed 25 ℃.
MSQ19 pure steam quality tester is designed to automatically monitor and calculate the content of non-condensable gases, dryness, and over-temperature based on requirements EN285 and HTM2010, displaying the results in real-time.
Fully-automated integrated design
no need to build the device, just connect the intake hose.
Quick detection
It only takes 5 minutes to complete the detection of three indicators, effectively avoid the security risks of manual operation and tedious data sorting and calculation.
Early warning function
Based on design EN285, the system automatically monitors the content of non-condensable gases, dryness, and over-temperature, and issues alarm when these values exceed the present limits.
Portable design
Features a removable design, which can not only meet the requirements of multi-point mobile detection, but also provide continuous data analysis for key positions.
Glove-friendly touching screen
Ergonomic large screen design, even when wearing gloves, the operation can be intuitive and smooth.
Data integrity
Authority management and audit tracking system, capable of storing at least 1000,000 groups of data.
Data printing
Integrated non-thermal printer for printing original data or exporting data in PDF format via USB interface.
Technical parameters of MSQ19 Fully-Automaticed Pure Steam Quality Tester
detection range | dryness | 80~100% |
non-condensable gas | 0~30% | |
overheat | ±50℃ | |
steam port size (external diameter) | TC25 chuck | |
electrical | voltage, frequency | 220V 50Hz |
parameter(complete machine) | power | 240W |
Order information
product number | M101 |
weight | 30kg |
size(LWH) | 510*342*645mm |